INSTRUCTION MANUAL
10 series
Blood Pressure Monitor
with ComFitTM Cuff
Model BP785
TABLE OF CONTENTS
Taking a Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Calibration Check System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Cuff Wrap Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Using the Guest Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Selecting the USER ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Using the USER ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Using the Single Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Using the TruReadTM Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
Special Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
Using the Memory Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Averaging Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
To Display the Measurement Values . . . . . . . . . . . . . . . . . . . . . . .37
Morning and Evening Averages . . . . . . . . . . . . . . . . . . . . . . . . . .39
Morning Averages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Evening Averages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
To Display Morning and Evening Averages . . . . . . . . . . . . . . . . .40
Display Combinations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
To Delete All Values Stored in the Memory . . . . . . . . . . . . . . . . .42
Care and Maintenance
Care and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
Error Indicators and Troubleshooting Tips . . . . . . . . . . . . . . . . . . . . . . . .45
FCC Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
Limited Warranties . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
3
Before Using the Monitor
INTRODUCTION
Thank you for purchasing the Omron® BP785 IntelliSense® Automatic
Blood Pressure Monitor with ComFitTM Cuff.
Fill in for future reference.
DATE PURCHASED: __________________
SERIAL NUMBER: ____________________
• Staple your purchase receipt here
• Register your product on-line at
Your new digital blood pressure monitor uses the oscillometric method
of blood pressure measurement. This means the monitor detects your
blood’s movement through your brachial artery and converts the
movements into a digital reading. The monitor stores measurement
results for two people and features Morning and Evening Averages.
TruReadTM allows for three consecutive measurements with timed
intervals for an average reading.
The BP785 comes with the following components:
• Monitor
• ComFitTM Cuff • AC Adapter
• Storage Case
• Instruction Manual printed in English and Spanish
• Quick Start Guide
The BP785 IntelliSense® Automatic Blood Pressure Monitor is intended
for home use.
SAVE THESE INSTRUCTIONS
4
SAFETY INFORMATION
To assure the correct use of the product basic safety measures should
always be followed including the warnings and cautions listed in this
instruction manual.
SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL
Indicates a potentially hazardous situation
which, if not avoided, could result in death or
serious injury.
WARNING
CAUTION
Indicates a potentially hazardous situation
which, if not avoided, may result in minor or
moderate injury to the user or patient or
damage to the equipment or other property.
OPERATING THE DEVICE
Contact your physician for specific information about your blood
pressure. Self-diagnosis and treatment using measured results may
be dangerous. Follow the instructions of your physician or licensed
healthcare provider.
DO NOT adjust medication based on measurement results
from this blood pressure monitor. Take medication as
prescribed by your physician. Only a physician is qualified to
diagnose and treat High Blood Pressure.
The monitor is not intended to be a diagnostic device.
If battery fluid should get in your eyes, immediately rinse with
plenty of clean water. Contact a physician immediately.
Read all of the information in the instruction manual and any other
literature in the box before operating the unit.
5
SAFETY INFORMATION
OPERATING THE DEVICE (continued)
This device is intended for use in measuring blood pressure and
pulse rate in the adult population. Do not use this device on infants
or persons who cannot express their intentions.
Read the Special Conditions section (page 34) of the instruction
manual if your systolic pressure is known to be more than 220
mmHg. Inflating to a higher pressure than necessary may result in
bruising where the cuff is applied.
Operate the device only as intended. Do not use the device for any
other purpose.
Dispose of the device, components and optional accessories
according to applicable local regulations. Unlawful disposal may
cause environmental pollution.
Do not use a cellular phone near the device. It may result in an
operational failure.
Use only Omron authorized parts and accessories. Parts and
accessories not approved for use with the device may damage the
unit.
Use only 1.5V alkaline batteries with this device.
Do not use other types of batteries. This may damage the unit.
6
SAFETY INFORMATION
RISK OF ELECTRICAL SHOCK
Do not plug or unplug the power cord into the electrical outlet with
wet hands.
Do not overload power outlets. Plug the device into the appropriate
voltage outlet.
CARE AND MAINTENANCE
Do not subject the monitor to strong shocks, such as dropping the
unit on the floor.
Do not submerge the device or any of the components in water.
Store the device and the components in a clean, safe location.
Changes or modification not approved by Omron Healthcare will
void the user warranty. Do not disassemble or attempt to repair the
unit or components.
7
BEFORE TAKING A MEASUREMENT
To ensure a reliable reading follow these recommendations:
1. Avoid eating, drinking alcohol, smoking, exercising, and bathing for
30 minutes prior to taking a measurement. Rest for at least 15
minutes prior to taking the measurement.
2. Stress raises blood pressure. Avoid taking measurements during
stressful times.
3. Measurements should be taken in a quiet place.
4. Remove tight-fitting clothing from your arm.
5. Sit on a chair with your feet flat on the floor. Rest your arm on
a table so that the cuff is at the same level as your heart.
6. Remain still and do not talk during the measurement.
7. Keep a record of your blood pressure and pulse readings for your
physician. A single measurement may not provide an accurate
indication of your true blood pressure. You need to take and record
several readings over a period of time. Try to measure your blood
pressure at about the same time each day for consistency.
8
Operating Instructions
KNOW YOUR UNIT
Main Unit:
Display
Calibration
Check System
SET Button
Morning/Evening Average
) Button
USER ID
Selection
Switch
(
Memory ( ) Button
Start/Stop Button
Up/Down
(
) Buttons
/
Battery Compartment
Air Jack
AC Adapter Jack
9
KNOW YOUR UNIT
Components:
Housing
Air Tube
Cord
AC Adapter
Air Plug
(Connects to Air Jack)
Arm Cuff
Instruction Manual
Storage Case
Quick Start Guide
10
UNIT DISPLAY
Morning Average Symbol
Evening Average Symbol
Date/Time Display
Irregular Heartbeat Symbol
USER ID Symbol (A or B)
Week Display
TruReadTM Symbol
Average Value Symbol
Memory Symbol
Systolic Blood Pressure
Diastolic Blood Pressure
Morning Hypertension Symbol
Movement Symbol
Blood Pressure Level Indicator
Deflation Symbol
Cuff Wrap Guide
Heartbeat Symbol
Low Battery Indicator
Pulse Display / Memory Number
11
DISPLAY SYMBOLS
IRREGULAR HEARTBEAT SYMBOL (
)
When the monitor detects an irregular rhythm two or more times during
the measurement, the Irregular Heartbeat Symbol ( ) will appear on the
display with the measurement values.
Normal Heartbeat
An irregular heartbeat rhythm is defined as
a rhythm that varies by less than 25% from
Pulse
the average rhythm or 25% more than the
average rhythm detected while the monitor
Blood
pressure
is measuring the systolic and diastolic blood
pressure.
Irregular Heartbeat
Short Long
Pulse
If the irregular heartbeat symbol (
)
displays with your measurement results, we
recommend you consult your physician.
Follow the directions of your doctor.
Blood
pressure
MOVEMENT ERROR SYMBOL (
)
The Movement Error Symbol is displayed if you move your body during
the measurement. Please remove the arm cuff, and wait 2-3 minutes.
Take another measurement, remain still during inflation.
USER ID SYMBOL (
/
)
The USER ID Symbol indicates USER A or USER B as selected by
the user to take a measurement or when using the memory function.
The USER ID Symbol is not displayed when using the Guest Mode to
take a measurement.
12
DISPLAY SYMBOLS
The American Heart Association1 recommends the following guideline
for upper limit of normal for home blood pressure readings:
AHA Home Guideline for Upper Limit of Normal BP
Systolic Blood Pressure
135 mmHg
Diastolic Blood Pressure
85 mmHg
This is a general guideline because blood pressure depends on
age and morbidity. The AHA recommends a lower home blood
pressure goal for certain patients, including diabetic patients,
pregnant women, and patients with renal failure1.
The monitor is not intended to be a diagnostic device.
MORNING HYPERTENSION SYMBOL (
)
The Morning Hypertension Symbol is displayed if the morning average
reading for a week is above 135 for the Systolic Blood Pressure value
and/or 85 for the Diastolic Blood Pressure value when viewing the
Morning Averages and the Evening Averages in the memory.
Omron Healthcare recommends you contact your physician if the
MORNING HYPERTENSION SYMBOL and/or the HEARTBEAT
SYMBOL appear on the display with your measurement values.
1American Heart Association Scientific Statement 2004
13
DISPLAY SYMBOLS
HEARTBEAT SYMBOL (
)
TAKING A MEASUREMENT
The Heartbeat Symbol flashes on the display during the
measurement. The Heartbeat Symbol flashes at every heartbeat.
AVERAGE VALUE SYMBOL (
)
The Average Value Symbol is displayed when you press the memory
button. The most recent average reading appears on the display screen.
MORNING AVERAGE SYMBOL (
)
The Morning Average symbol is displayed when viewing Morning
Averages using the Memory Function.
EVENING AVERAGE SYMBOL (
)
The Evening Average symbol is displayed when viewing Evening
Averages using the Memory Function.
CUFF WRAP GUIDE SYMBOL(
/
)
When the cuff is applied correctly, ( ) is displayed while taking a
measurement or using the memory function. If the cuff is not applied
correctly, (
) is displayed.
14
USING THE AC ADAPTER
IMPORTANT SAFETY INSTRUCTIONS
SAVE THESE INSTRUCTIONS
- DANGER -
TO REDUCE THE RISK OF
FIRE OR ELECTRIC SHOCK,
CAREFULLY FOLLOW THESE INSTRUCTIONS.
For connection to a power supply not in the U.S.A., use an
attachment plug adapter of the proper configuration for the power
outlet.
The power unit (AC Adapter) is intended to be correctly oriented
in a vertical or floor mount position.
NOTES: • Use only the authorized Omron AC Adapter that came with
this monitor.
• We recommend you install the batteries even if the AC
Adapter is used. If no batteries are installed, you may need to
reset the date and time if the AC Adapter is disconnected.
The measurement results will not be deleted.
CAUTION
Use only Omron authorized parts and accessories. Parts and
accessories not approved for use with the device may damage the
unit.
15
USING THE AC ADAPTER
TO CONNECT THE AC ADAPTER
① Insert the AC Adapter Plug into
the AC Adapter Jack on the back
side of the monitor as illustrated.
Housing
② Plug the AC Adapter into a 120V
AC outlet (60 cycles).
Cord
TO DISCONNECT THE AC ADAPTER
① Hold the Housing to remove the AC Adapter from the electrical
outlet.
② Remove the AC Adapter plug from the monitor.
WARNING
Do not plug or unplug the power cord into the electrical outlet with
wet hands.
CAUTION
Do not overload power outlets. Plug the device into the appropriate
voltage outlet.
16
BATTERY INSTALLATION
Battery Cover
1. Remove the battery cover with pulling the
hook up while pushing it in the direction
of the arrow.
2. Install 4 “AA” size batteries so the +
(positive) and - (negative) polarities match
the polarities of the battery compartment
as indicated.
3. Replace the battery cover.
17
BATTERY INSTALLATION
BATTERY REPLACEMENT
Low Battery Indicator
When the Low Battery Indicator appears on the display screen turn the
monitor off and remove all the batteries. Replace with four new batteries
at the same time. Longlife alkaline batteries are recommended.
CAUTION
If battery fluid should get in your eyes, immediately rinse with
plenty of clean water. Contact a physician immediately.
NOTES: • Turn the unit off before replacing the batteries. If the
batteries are removed while the unit is still on, the date and
time will be reset to that of the previous use. However, the
measurement results are not deleted.
• When the batteries are replaced, you may need to reset the
date and time. If “-:--” appears on the display, refer to page
19, SETTING THE DATE/TIME AND TRUREADTM.
• The supplied batteries may have a shorter life.
CAUTION
Dispose of the device, components and optional accessories
according to applicable local regulations. Unlawful disposal may
cause environmental pollution.
18
SETTING THE DATE/TIME AND TRUREAD TM
Set the monitor to the current date and time before taking a
measurement for the first time. If the date and time are not set correctly,
the measurement values stored in the memory and the average values
may not be correct.
If the date and time are not set, “-:--” appears during or after
measurement.
The Blood Pressure Monitor automatically stores up to 100 individual
measurement values with the dates and times. The monitor also
calculates weekly averages for measurements in the morning and in the
evening. Morning and Evening averages are stored for this week and the
last seven weeks.
The monitor can also calculate an average reading based on the three
most recent measurements taken within 10 minutes of the last reading.
NOTE: • If only two readings are stored in the memory for the 10-
minute period, the average is based on the two readings.
• If one reading is stored in the memory for the 10-minute
period, this reading is displayed as the average.
19
SETTING THE DATE/TIME AND TRUREAD TM
1. SETTING THE TRUREADTM MODE
1) Select your USER ID (A or B).
2) Press the SET button.
The TruReadTM symbol
on the display.
and the oFF setting appears
3) Press the Up or Down button (
) to select on.
/
NOTE: Press the Up or Down
button to alternate between
on and oFF.
20
SETTING THE DATE/TIME AND TRUREAD TM
4) Press the SET button to confirm the change.
If on is selected, the number of seconds between each
measurement is displayed.
NOTE: The default value between each measurement is 60
seconds.
5) Press the Up or Down (
between each measurement.
) to change the interval of time
/
The interval can be set to 15 seconds, 30 seconds, 60 seconds or
120 seconds. Continue to press the Up or Down button until the
desired interval appears on the display.
6) Press the SET button to confirm the setting.
The year flashes on the display.
21
SETTING THE DATE/TIME AND TRUREAD TM
2. SETTING THE YEAR
The year can be set between 2010 and
2040. When the display reaches 2040,
it will return to 2010.
Press the Up button (
) to advance
by increments of one year.
Press the SET button to set the current
year.
The month flashes on the display.
NOTES: • Press the Down button (
) to decrease the date and time
values by decrements of one.
• Press and hold the Up or Down button to increase or
decrease the date and time values faster.
3. SETTING THE MONTH
Press the Up button ( ) to advance
by increments of one month.
Press the SET button to set the current
month.
The day flashes on the display.
22
SETTING THE DATE/TIME AND TRUREAD TM
4. SETTING THE DAY
Press the Up button (
) to advance
by increments of one day.
Press the SET button to set the current
day.
The hour flashes on the display.
5. SETTING THE HOUR
The time is set using AM or PM.
Press the Up button (
) to advance
by increments of one hour.
Press the SET button to set the current
hour.
The minute flashes on the display.
6. SETTING THE MINUTE
Press the Up button (
) to advance by increments of one minute.
Press the SET button to set the minute setting.
7. TO TURN THE MONITOR OFF
Press the START/STOP button.
23
APPLYING THE ARM CUFF
CAUTION
Read all of the information in the instruction manual and any other
literature in the box before operating the unit.
CAUTION
This device is intended for use in measuring blood pressure and
pulse rate in the adult population. Do not use this device on infants
or persons who cannot express their intentions.
APPLYING THE CUFF ON THE LEFT ARM
1. Make sure the air plug is securely inserted in the main unit.
2. Remove tight-fitting clothing from your upper left arm.
24
APPLYING THE ARM CUFF
3. Sit on a chair with your feet flat on the floor. Place your left arm
on a table so the cuff is level with your heart.
Thumb Grip
4. Hold the thumb grip on the cuff
securely with your right hand.
Cloth Fastener
5. Turn the palm of your left hand upward.
25
APPLYING THE ARM CUFF
6. Apply the cuff to your left upper arm so the blue stripe is on the
inside of your arm and aligned with your middle finger. The air tube
runs down the inside of your arm. The bottom of the cuff should be
approximately 1/2" above your elbow.
Blue Stripe
1/2 inch (1 - 2 cm)
Air Tube
7. Wrap the cuff firmly in place around your arm using the cloth
fastener.
26
APPLYING THE ARM CUFF
APPLYING THE CUFF ON THE RIGHT ARM
When taking a measurement using the right arm use this instruction
for Step 6 on page 26.
Apply the cuff to your right upper arm so the thumb grip is centered
on the inside of your inner arm. The bottom of the cuff should be
approximately 1/2'' above your elbow.
Thumb Grip
NOTE: Be careful not to rest your arm on the air tube. This will restrict
the flow of air to the cuff.
27
TAKING A MEASUREMENT
The monitor is designed to take measurements and store the
measurement values in the memory for two people using
USER ID A and USER ID B.
The monitor can also be used to take a single measurement for other
persons using the GUEST Mode. The USER ID symbol is not displayed
when using the GUEST Mode. Measurements taken using the GUEST
Mode are not stored in the memory.
WARNING
Contact your physician for specific information about your blood
pressure. Self-diagnosis and treatment using measured results may
be dangerous. Follow the instructions of your physician or licensed
healthcare provider.
CAUTION
Operate the device only as intended. Do not use the device for any
other purpose.
CALIBRATION CHECK SYSTEM
When you press the START/STOP button, the Calibration Check
System light turns on and the unit starts monitoring your readings
using dual sensors. If the unit is accurate and functioning correctly,
the Calibration Check System light remains lit during the
measurement. If an error is detected, the Calibration Check System
light flashes and “ER” will appear on the display.
CUFF WRAP GUIDE
This monitor checks whether the arm cuff is applied correctly during the
inflation. When the cuff is applied correctly,
taking a measurement or using the memory function. If the cuff is not
applied correctly, is displayed. If is displayed, refer to the
is displayed while
section “Applying the Arm Cuff”, reapply the cuff and take another
measurement.
28
TAKING A MEASUREMENT
USING THE GUEST MODE
1. Press and hold the START/STOP button.
The A and B Symbols disappear from the display.
A and B Symbols disappear
2. Release the START/STOP button.
The cuff starts to inflate automatically.
NOTE: To stop the inflation press the START/STOP button.
The monitor will stop inflating, start deflation and turn off.
3. Inflation stops and the measurement is started. When the
measurement is completed, your blood pressure and pulse rate appear
on the display.
29
TAKING A MEASUREMENT
SELECTING THE USER ID
Select your USER ID. Always use the same USER ID when taking a
measurement. The monitor stores the measurement values in the
memory by the USER ID selected. These measurement values are used
to calculate the Morning Averages and Evening Averages.
Slide the USER ID Selection Switch right or left to select USER A or B.
USING THE USER ID
When you select a USER ID the monitor can be set to take one
measurement using the Single Mode or three consecutive measurements
using the TruReadTM Mode when you press the START/STOP button.
NOTE: The default value is set to the Single Mode to take one
measurement.
The TruReadTM Mode takes three consecutive measurements.
The monitor will inflate, take a measurement, and deflate three times
separated by a short interval of time between each measurement.
30
TAKING A MEASUREMENT
USING THE SINGLE MODE
1. Select your USER ID (A or B).
2. Press the START/STOP button.
All symbols appear on the display. The USER ID Symbol you
selected flashes on the display.
The cuff starts to inflate automatically. As the cuff inflates, the monitor
automatically determines your ideal inflation level.
This monitor detects the pulse during inflation. Do not move your arm
and remain still until the entire measurement process is completed.
NOTE: To stop the inflation or measurement, press the START/STOP
button. The monitor will stop inflating, start deflating and will
turn off.
31
TAKING A MEASUREMENT
3. Inflation stops automatically and the measurement is started.
As the cuff deflates, decreasing numbers appear on the display.
The Heartbeat Symbol ( ) flashes at every heartbeat.
4. When the measurement is complete, the arm cuff
completely deflates. Your blood pressure and pulse
rate are displayed.
5. Press the START/STOP button to turn the
monitor off.
NOTE: The monitor will automatically turn
off after two minutes.
32
TAKING A MEASUREMENT
USING THE TRUREADTM MODE
1. Select your USER ID (A or B).
2. Press the START/STOP button.
All symbols appear on the display. The USER ID Symbol you
selected flashes on the display. The cuff starts to inflate automatically.
NOTE: To stop the inflation or measurement, press the START/STOP
button. The monitor will stop inflating, start deflating and will
turn off.
3. Inflation stops automatically and the measurement is started.
As the cuff deflates, decreasing numbers appear on the display.
The Heartbeat Symbol ( ) flashes at every heartbeat. When the
measurement is complete, the arm cuff completely deflates.
NOTE: Do not move your arm and remain still until the entire
measurement process is completed.
4. Check the progress of the consecutive measurements.
NOTE: If a measurement fails, the monitor will repeat taking
a measurement up to five times during the consecutive
measurement process.
The first measurement is complete, waiting to start the next
measurement. The interval of time set between each measurement will
appear on the display.
33
TAKING A MEASUREMENT
The second measurement is complete, waiting to start the
next measurement.
The consecutive measurements are complete. The average for three
measurements appears on the display.
5. Press the SET button to display the blood pressure and pulse rate for
the first measurement. Press the SET button again to display the next
measurement.
6. Press the START/STOP button to turn the monitor off.
WARNING
Contact your physician for specific information about your blood
pressure. Self-diagnosis and treatment using measured results may
be dangerous. Follow the instructions of your physician or licensed
healthcare provider.
WARNING
DO NOT adjust medication based on measurement results from this
blood pressure monitor. Take medication as prescribed by your
physician. Only a physician is qualified to diagnose and treat High
Blood Pressure.
34
TAKING A MEASUREMENT
WARNING
The monitor is not intended to be a diagnostic device.
SPECIAL CONDITIONS
CAUTION
Inflating to a higher pressure than necessary may result in bruising
where the cuff is applied.
If your systolic pressure is known to be more than 220 mmHg, press
and hold the START/STOP button until the monitor inflates 30 to 40
mmHg higher than your expected systolic pressure.
NOTES: • Do not apply more pressure than necessary.
• The monitor will not inflate above 299 mmHg.
When the measurement is complete, the arm cuff completely deflates.
Your blood pressure and pulse rate are displayed.
35
USING THE MEMORY FUNCTION
The monitor is designed to store the blood pressure and the pulse
rate in the memory for two people (USER A and USER B) every time
a measurement is completed.
The monitor automatically stores up to 100 sets of measurement values
(blood pressure and pulse rate) for each user (A and B). When 100 sets
of measurement values are stored, the oldest record is deleted to save
the most recent values. The monitor also stores 8 weeks of Morning
Averages and 8 weeks of Evening Averages for each user (A and B).
AVERAGING FUNCTION
The monitor calculates the average reading based on the three most
recent sets of measurement values taken within 10 minutes of the most
recent reading.
NOTES: • If two sets of measurement values are stored in the memory
for the 10 minutes period, the average is based on the two
sets of measurement values. If one set of measurement
values is stored, this is displayed as the average.
• When viewing the reading taken without setting the date and
time, “-:--” is displayed instead of the date and time.
36
USING THE MEMORY FUNCTION
TO DISPLAY THE MEASUREMENT VALUES
1. Select your USER ID (A or B).
2. Press the Memory button ( ).
The Average Symbol ( ) appears on the display when the average
reading is shown.
37
USING THE MEMORY FUNCTION
3. Press the Down button (
) to display the most recent set of
measurement values on the screen. The values are displayed from the
most recent to the oldest.
NOTE: The date and time are alternately
displayed with the measurement
values.
Alternating display
Press the Up or Down button to display the
next set of values.
: To the older readings
: To the more recent readings
NOTE: Press and hold the Up or Down button to display the values
faster.
TruReadTM Mode
The average for the consecutive measurements appears on the display
with the TruReadTM symbol (
). To display the values for the
individual measurements press the SET button while the average is
displayed.
4. Press the START/STOP button to turn the power off.
38
USING THE MEMORY FUNCTION
MORNING AND EVENING AVERAGES
The monitor calculates and displays a weekly average for measurements
taken in the morning ( ) and evening ( ). The monitor stores 8
weeks of Morning Averages and 8 weeks of Evening Averages for each
user (A and B).
NOTE: The week begins Sunday at 4:00 AM
MORNING AVERAGES
Morning averages are based on the first average reading using the
TruReadTM Mode or the first reading using the Single Mode.
Measurement times must be between:
4:00 AM and 9:59 AM.
EVENING AVERAGES
Evening averages are based on the last average reading using the
TruReadTM Mode or the last reading using the Single Mode.
Measurement times must be between:
7:00 PM and 1:59 AM.
39
USING THE MEMORY FUNCTION
TO DISPLAY MORNING AND EVENING AVERAGES
1. Select your USER ID (A or B).
2. Press the Morning/ Evening
Average button (
) to select
the Morning or Evening average.
The average for the current week
“THIS WEEK” appears on
the display.
3. Press the Down (
) button to display the previous weeks.
With each press of the Down (
from 1 week ago to the older.
With each press of the Up (
7 week ago to the recent.
) button, the values are displayed
) button, the values are displayed from
NOTE: If no measurement is taken in a given week, the
display will appear as illustrated.
4. Press the START/STOP button to turn the monitor off.
40
USING THE MEMORY FUNCTION
DISPLAY COMBINATIONS
In addition to the Morning Average and Evening Average Symbols, the
monitor may also display the Morning Hypertension Symbol if the
morning average for that week is above the AHA guidelines. (See page
13 for details.) Depending on your measurement results, these may be
displayed in the following combinations.
Display
Morning
Average
Morning average from 7 weeks ago
Morning average from This Week
with Morning Average Symbol +
with Morning Average Symbol
Morning Hypertension Symbol
Evening
Average
Evening average from 7 weeks ago
with Evening Average Symbol +
Morning Hypertension Symbol
Evening average from This Week
with Evening Average Symbol
Average above AHA guidelines +
Morning Hypertension
Average within AHA guidelines
41
USING THE MEMORY FUNCTION
The Morning Hypertension Symbol ( ) appears if the weekly average
for morning measurements is above 135/85.
In this case the Morning Hypertension Symbol ( ) is displayed when
the Evening Average is displayed, regardless of the values for the
Evening Average.
TO DELETE ALL VALUES STORED IN THE MEMORY
The values stored in the memory are deleted by USER ID.
You cannot partially delete values stored in the memory. All values for
the user you select will be deleted.
1. Select your USER ID (A or B).
2. While holding the Memory button ( ) and the START/STOP button
simultaneously for more than 2 seconds all values will be deleted.
42
Care and Maintenance
CARE AND MAINTENANCE
To keep your digital blood pressure monitor in the best condition and
protect the unit from damage, follow the directions listed below:
Keep the monitor in the storage case
when not in use. Make sure the AC
adapter is placed under the main unit so
that it does not damage the display.
Avoid kinking or sharply bending the AC
Adapter cord.
Do not forcefully bend the arm cuff or
air tube. Do not fold tightly.
Clean the monitor with a soft dry cloth. Do not use any abrasive
or volatile cleaners. Do not attempt to clean the cuff.
CAUTION
Do not submerge the device or any of the components in water.
Do not subject the monitor to extreme hot or cold temperatures,
humidity or direct sunlight.
CAUTION
Store the device and the components in a clean, safe location.
CAUTION
Do not subject the monitor to strong shocks, such as dropping the
unit on the floor.
Remove the batteries if the unit will not be used for three months or
longer. Always replace all the batteries with new ones at the same time.
43
CARE AND MAINTENANCE
Use the unit consistent with the instruction provided in this manual.
CAUTION
Changes or modification not approved by Omron Healthcare will
void the user warranty. Do not disassemble or attempt to repair the
unit or components.
CAUTION
Use only Omron authorized parts and accessories. Parts and
accessories not approved for use with the device may damage
the unit.
44
ERROR INDICATORS AND TROUBLESHOOTING TIPS
ERROR INDICATORS
SYMBOL
CAUSE
CORRECTION
Insert the plug securely. Refer to “Applying the
Arm Cuff”.
Air plug disconnected.
Apply the arm cuff correctly. Refer to “Applying
the Arm Cuff”.
Arm cuff not applied correctly.
Air is leaking from the arm cuff.
Replace the cuff with the new one.
Repeat measurement. Remain still and do not talk
during measurement. Refer to “Taking a
Measurement”.
Cuff over inflated.
If “E2” appears repeatedly, inflate the cuff manually
until it is 30 to 40 mmHg above your previous
measurement result. Refer to “Taking a Measurement”.
The arm cuff was inflated above 299
Do not inflate the cuff above 299 mmHg. Refer to
mmHg when inflating the cuff manually. “Taking a Measurement”.
Repeat measurement. Remain still and do not talk
during measurement. Refer to “Taking a
Measurement”.
Movement during measurement.
Clothing is interfering with the
arm cuff.
Remove any clothing interfering with the arm cuff.
Refer to “Applying the Arm Cuff”.
Device error.
Contact Customer Service.
Sensor may be out of alignment.
Contact Customer Service for repair information.
Cuff is not applied correctly.
Refer to “Applying the Arm Cuff”.
45
ERROR INDICATORS AND TROUBLESHOOTING TIPS
ERROR INDICATORS
SYMBOL
CAUSE
CORRECTION
Replace the four batteries.
Refer to “Battery Installation.”
Batteries are worn
TROUBLESHOOTING TIPS
PROBLEM
CAUSES AND SOLUTIONS
No power.
Replace all four batteries with new
ones.
No display appears on the unit. Check the battery installation
for proper placement of the
battery polarities.
Blood pressure varies constantly.
Many factors including stress, time
of day, and how you wrap the cuff,
may affect your blood pressure.
Review the sections “Before Taking
Measurement values appear
too high or too low.
a Measurement” and “Taking a
Measurement.”
46
FCC STATEMENT
NOTE:
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)
This product has been tested and found to comply with the limits for a Class B
digital device, pursuant to part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful
interference in a residential installation. The product generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance
with the instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular
installation. If the product does cause harmful interference to radio or television
reception, which can be determined by turning the product on and off, the
user is encouraged to try to correct the interference by one or more of the
following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the product and the receiver.
• Connect the product into an outlet on a circuit different from that to which
the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only)
This digital apparatus does not exceed the Class B limits for radio noise
emissions from digital apparatus as set out in the interference-causing equipment
standard entitled “Digital Apparatus”, ICES-003 of the Canadian Department of
Communications.
Cet appareil numérique respecte les limites de bruits radioeléctriques applicables
aux appareils numériques de Clase B prescrites dans la norme sur le materiel
brouilleur: “Appareils Numériques”, ICES-003 édictée par le minister des
communications.
Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
47
LIMITED WARRANTIES
Your Omron® BP785 IntelliSense® Automatic Blood Pressure Monitor, excluding
the arm cuff, is warranted to be free from defects in materials and workmanship
appearing within 5 years from the date of purchase, when used in accordance
with the instructions provided with the monitor. The arm cuff is warranted to be
free from defects in materials and workmanship appearing within one year from
the date of purchase when the monitor is used in accordance with the instructions
provided with the monitor. The above warranties extend only to the original retail
purchaser.
We will, at our option, repair or replace without charge any monitor or arm cuff
covered by the above warranties. Repair or replacement is our only responsibility
and your only remedy under the above warranties.
To obtain warranty service contact Customer Service by calling 1-800-634-4350 for
the address of the repair location and the return shipping and handling fee.
Enclose the Proof of Purchase. Include a letter, with your name, address, phone
number, and description of the specific problem. Pack the product carefully to
prevent damage in transit. Because of possible loss in transit, we recommend
insuring the product with return receipt requested.
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON
IN CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY
DISCLAIMS ANY OTHER WARRANTIES, EXPRESS OR IMPLIED,
INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITY AND
FITNESS FOR A PARTICULAR PURPOSE. IMPLIED WARRANTIES
AND OTHER TERMS THAT MAY BE IMPOSED BY LAW, IF ANY, ARE
LIMITED IN DURATION TO THE PERIOD OF THE ABOVE EXPRESS
WARRANTY.
OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER
SPECIAL, INCIDENTAL, CONSEQUENTAIL OR INDIRECT COSTS,
EXPENSES OR DAMAGES.
This warranty provides you with specific legal rights, and you may have other
rights that vary by jurisdiction. Because of special local requirements, some of
the above limitations and exclusions may not apply to you.
FOR CUSTOMER SERVICE
Visit our web site at:
Call toll free:
1-800-634-4350
48
SPECIFICATIONS
Model:
BP785 (HEM-7222-Z)
Display:
LCD Digital Display
Measurement Range: Pressure: 0 to 299 mmHg, Pulse: 40 to 180/min.
Accuracy/Calibration: Pressure: 3mmHg or 2% of reading
Pulse:
5% of reading
Inflation:
Deflation:
Rapid Air Release:
Automatic by electric pump
Automatic pressure release valve
Automatic exhaust valve
Measurement Method: Oscillometric method
Power Source:
Battery Life:
120V 60Hz AC adapter or 4 “AA” batteries
Approx. 500 uses when used 6 times a day
with 4 new alkaline batteries
Operating Temperature 50ºF to 104ºF (10ºC to 40ºC) /15 to 90% RH
/Humidity:
Storage Temperature
/Humidity/Air Pressure: /700 - 1060 hPa
Main Unit Weight: Approximately 15 1/8 oz. (430 g)
not including batteries
-4ºF to 140ºF (-20ºC to 60ºC) /10 to 95% RH
Main Unit Dimensions: Approx. 6 1/4" (l) × 4 7/8" (w) × 3 5/16" (h)
(158 mm × 124 mm × 86 mm)
Cuff Size:
Approximately 6" (w) (152 mm)
Cuff tube 23 5/8" (l) (600 mm)
Fits arm circumferences 9" to 17"
(220 mm to 420 mm)
Cuff Circumference:
Memory:
Contents:
Up to 100 per user
Main Unit, ComFitTM Cuff, AC Adapter,
Instruction Manual, Quick Start Guide, Storage Case
NOTE: These specifications are subject to change without notice.