OPERATION:  
CALIBRATION PROCEDURE:  
Milwaukee  
The meter is supplied complete with a 9V  
battery. Slide off the battery compartment  
cover on the back of the meter. Install the  
battery into the battery clip connector while  
observing polarity.  
A) Preparation:  
USER MANUAL  
Two calibration buffers are required  
1. pH 7.01 (MA9007)  
2. pH 4.01 (MA9004) if you are measuring  
in acid range (pH 0 - pH 7) or pH 10.01  
(MA9010) if you are measuring in  
alkaline range (pH 7 - pH 14).  
PORTABLE pH METER  
MODEL: MW101  
Always remove the electrode protective  
cap before taking any measurement. If  
the electrode has been left dry, soak the  
tip (bottom 2.5 cm) in rinse solution  
(M10000B) for a few minutes to reactivate  
it.  
Use two beakers for each pH buffer. One  
beaker for rinsing the electrode, the other  
for calibration.  
Smart pH Meter  
Use a thermometer with 1ºC accuracy to  
measure the temperature of calibration  
solution.  
Connect the pH electrode  
to the BNC socket on  
the top of the meter.  
B) Procedure:  
Remove the protective cap  
from the electrode. Rinse  
the tip of the electrode with  
some pH 7.01 solution, then  
immerse the pH electrode  
into a pH 7.01 buffer solution.  
Turn the instrument on  
by pressing the ON/OFF  
key.  
Make sure that the meter has been  
calibrated before taking any measurements  
(see Calibration Procedure).  
Take the temperature of  
the buffer solution with a  
thermometer and set the  
temperature knob to the  
measured temperature (e.g.  
15ºC).  
Set the temperature knob  
to the value of testing  
solution  
with help of an accurate  
thermometer).  
(measured  
WARRANTY:  
This instrument is warranted from all defects in materials  
and manufacturing for a period of two years from the  
date of purchase.  
Adjust the OFFSET trimmer  
(pH 7) on the front panel, with  
a small screwdriver until the  
LCD shows the pH value at  
temperature of the buffer (see  
the pH versus temperature  
chart).  
Immerse the tip (2.5 cm)  
of the pH electrode into  
the sample and stir  
gently.  
The electrode is warranted for a period of six  
months.  
If during this period the repair or replacement of parts  
is required, where the damage is not due to negligence  
or erroneous operation by the user, please return the  
parts to either distributor or our office and the repair will  
be effected free of charge.  
After completing measurements,  
switch the meter off and store  
the electrode with a few drops  
of storage solution (MA9015)  
in the protective cap.  
E.g. in this case, if the  
temperature is 15ºC, the  
meter display should be  
adjusted to read “pH 7.04”.  
Note: We reserve the right to modify the design,  
construction and appearance of our products  
without advance notice.  
Now rinse the pH electrode  
in the first pH 4.01 beaker,  
then immerse it into the  
second pH 4.01 beaker or  
follow the same procedure  
if using pH 10.01 buffer.  
Milwaukee  
 

Miele Cooktop KM 391 User Manual
Miller Electric Welder 500SSX User Manual
National Instruments Network Card NI 9426 User Manual
NEC Projector M260W and M300W User Manual
Nokia Car Stereo System 3586I User Manual
Olympus Camcorder C 575 User Manual
Optoma Technology Home Theater Screen DS 9092PM User Manual
Oregon Scientific Digital Camera DS9320 15 User Manual
Oster Convection Oven OTM1101VBS User Manual
Panasonic DVD Player DVD F84 User Manual