OPERATION:
CALIBRATION PROCEDURE:
Milwaukee
•
The meter is supplied complete with a 9V
battery. Slide off the battery compartment
cover on the back of the meter. Install the
battery into the battery clip connector while
observing polarity.
A) Preparation:
USER MANUAL
Two calibration buffers are required
1. pH 7.01 (MA9007)
2. pH 4.01 (MA9004) if you are measuring
in acid range (pH 0 - pH 7) or pH 10.01
(MA9010) if you are measuring in
alkaline range (pH 7 - pH 14).
PORTABLE pH METER
MODEL: MW101
•
Always remove the electrode protective
cap before taking any measurement. If
the electrode has been left dry, soak the
tip (bottom 2.5 cm) in rinse solution
(M10000B) for a few minutes to reactivate
it.
Use two beakers for each pH buffer. One
beaker for rinsing the electrode, the other
for calibration.
Smart pH Meter
Use a thermometer with 1ºC accuracy to
measure the temperature of calibration
solution.
•
•
•
•
Connect the pH electrode
to the BNC socket on
the top of the meter.
B) Procedure:
•
•
•
Remove the protective cap
from the electrode. Rinse
the tip of the electrode with
some pH 7.01 solution, then
immerse the pH electrode
into a pH 7.01 buffer solution.
Turn the instrument on
by pressing the ON/OFF
key.
Make sure that the meter has been
calibrated before taking any measurements
(see Calibration Procedure).
Take the temperature of
the buffer solution with a
thermometer and set the
temperature knob to the
measured temperature (e.g.
15ºC).
Set the temperature knob
to the value of testing
solution
with help of an accurate
thermometer).
(measured
WARRANTY:
This instrument is warranted from all defects in materials
and manufacturing for a period of two years from the
date of purchase.
Adjust the OFFSET trimmer
(pH 7) on the front panel, with
a small screwdriver until the
LCD shows the pH value at
temperature of the buffer (see
the pH versus temperature
chart).
•
•
Immerse the tip (2.5 cm)
of the pH electrode into
the sample and stir
gently.
The electrode is warranted for a period of six
months.
If during this period the repair or replacement of parts
is required, where the damage is not due to negligence
or erroneous operation by the user, please return the
parts to either distributor or our office and the repair will
be effected free of charge.
After completing measurements,
switch the meter off and store
the electrode with a few drops
of storage solution (MA9015)
in the protective cap.
E.g. in this case, if the
temperature is 15ºC, the
meter display should be
adjusted to read “pH 7.04”.
Note: We reserve the right to modify the design,
construction and appearance of our products
without advance notice.
•
Now rinse the pH electrode
in the first pH 4.01 beaker,
then immerse it into the
second pH 4.01 beaker or
follow the same procedure
if using pH 10.01 buffer.
Milwaukee
|